Terminated

ConMed Stealth coated Laparoscopic Electrodes with 4mm Pin (60-5158 series). Cat. #60-5158-034 - Spatula; Cat. #60-5158-234 -''J'' Hook; Cat. #60-5158-432 - ''L'' Hook; Cat. # 60-5158-934 - Needle. FIRM ON LABEL: ConMed Corporation, Utica, New York, USA 13502.

FDA Recall Z-1301-04 · Posted 2004-07-31

Recalling FirmConmed Corporation
Recall StatusTerminated
Product CodeGEI
Date Initiated2004-06-28
Date Terminated2005-10-05
Product Quantity3,252 units

Reason for Recall

Insulation of some electrodes may have pulled away from the hub of the electrode exposing the conductive shaft below. There is potential for electrical shock.

Firm Action

Letters dated 6/28/04 via Priority Mail with instructions to return product and to subrecall if further distributed.

Root Cause

Other

Distribution Pattern

Nationwide distribution to 24 hospitals, 3 distributors and 13 sales reps. Six foreign consignees in Mexico, Italy, United Kingdom, Spain, Turkey and Australia.