Presource Procedure Packs containing AAMI Level 3 surgical gowns, Multiple Catalog Numbers Produced Between September 1, 2018 and January 15, 2020
FDA Recall Z-1301-2020 · Posted 2020-02-04
Advertisement
Recalling FirmCardinal Health 200, LLC
Recall StatusTerminated
Product CodeLRO
Date Initiated2020-01-31
Date Terminated2023-02-01
Product Quantity318,000 units
Reason for Recall
The kits contained recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no assurance that the surgical gowns are sterile. An inadequately sterilized surgical gown could compromise a sterile field and increase the risk of a surgical site infection.
Firm Action
Urgent Medical Device Correction notification letters dated 1/31/20 were distributed to customers.
Actions Required:
1. CHECK all storage and usage locations to confirm if you have any units of the affected packs and lot
numbers per enclosed report identifying your facilitys packs.
2. ALERT your clinicians of this action including Exhibit A and B.
3. SEGREGATE and QUARANTINE all Presource Procedure Packs on-hand that are confirmed to be
affected by lot number, if not already completed, per enclosed report identifying your facilitys packs.
4. AFFIX a WARNING LABEL as outlined in Exhibit A and included with this notice to the front of each
Presource pack so that it is clearly visible to clinicians, instructing them to remove and discard the
recalled component(s) as well as any other componentry NOT contained within an inner, sealed
packaging separating the product from contacting the affected gown. All components within inner,
sealed packaging or other packs within the sterilization pouch may be used.
Refer to Exhibit B for detailed instructions.
Note: Additional labels can be sent to your facility upon request. Please email GMBCardinalSurgicalGownRecall@
cardinalhealth.com for assistance.
5. RETURN the enclosed acknowledgment form either by facsimile (614-495-5651) or email (GMBCardinalSurgicalGownRecall@
cardinalhealth.com) confirming your receipt of this action and the
completion of the labeling product. Please respond even if you are not affected by this medical device
correction notice.
6. NOTIFY any customers to whom you may have distributed or forwarded product affected by this
medical device correction notice. You may include a copy of this recall notice with your customer
notification.
7. CONTACT Presource Sales Operations at (800) 766-0706 or your sales representative if you do not
wish to accept overlabeled product or overlabel the product in your possession. For component credit
or replacement, please contact the followi
Root Cause
Other
Distribution Pattern
US Nationwide. Chile, Israel, UAE.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.