Recall StatusOpen, Classified
Date Initiated2022-05-13
Product Quantity7,520 units
Reason for Recall
Digital IVUS catheter instructions for use ( IFU) do not contain contrainidications
Firm Action
On May 25, 2022, Philips issued a "Urgent Medical Device Correction" notification. Phillips asked customers to take the following actions:
1. Philips recommends reviewing the enclosed compliant Instructions for Use, in Appendix A, before proceeding as normal.
2. As per standard practice, precaution should always be used when advancing or removing a catheter in complex vessel anatomies, such as not forcing a catheter into a narrow vessel or a tight stenosis. Vessel calcification, tortuosity and pre-existing untreated vessel spasm are key components to any complex vascular anatomy. Additionally, if the initial guidewire faces resistance or requires further manipulations, please use this as an indicator of complex patient anatomy.
3. Philips is providing this Urgent Medical Device Correction letter to affected customers and recommends forwarding this notification to all staff who need to be aware within your organization or to any organization where the potentially affected devices may have been transferred.
4. To acknowledge receipt of this notification, please complete, sign, and return the Customer Reply Form within 30 days upon receipt of this notice to Email:
[email protected]
Root Cause
Process control
Distribution Pattern
Worldwide distribution - US Nationwide distribution in the states of NY, CA, AZ, LA, NV, AL, MI, WI, CO, TX, NJ, OR, FL, KS, IL, MD, PA, OH, NC, OK, MN, AK, GA, MO, SC, MS, KY, AR, VA, SD, TN, MA, DE, NE, MT, CT, WV, WA, IN, IA, HI, NH, WY, UT, ID, DC, RI, NM, ME, VT and the countries of ES, SE, FR, PR, IT, DE, CA, NZ, PL, BR, ZA, NL.