Optiflux 200NRe Dialyzer Finished Assy., Model no. 0500320E
FDA Recall Z-1303-2022 · Posted None
Advertisement
Recalling FirmFresenius Medical Care Holdings, Inc.
Recall StatusTerminated
Product CodeKDI
Date Initiated2022-05-04
Date Terminated2024-05-06
Product Quantity19 cases
Reason for Recall
Dialyzers were distributed after being exposed to cold temperatures which exceeded the low-temperature range in the firm's standard operating procedure.
Firm Action
A Customer Notification was sent to the five (5) affected consignees on 05/04/2022 via USPS First Class Mail. The firm also followed up with phone calls with consignees with instructions to examine their stock to determine whether they have affected product on hand, and to remove product from their inventory.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution in the state of Pennsylvania.