Recall StatusOpen, Classified
Date Initiated2024-02-08
Product Quantity64,009 kits
Reason for Recall
Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results.
Firm Action
The recalling firm issued letters on 2/8/2024 via email and standard postal mail to their U.S. customers and postal mail to their OUS customers. The letter explained the issue, impact, and actions to be taken. The actions included: (1) Discontinue using the affected lots and discard all remaining reagent packs from the lots; (2) The firm recommended performing a retrospective review on patient results generated with the affected reagent lots and assess the need for retesting;(3) Contact the local Beckman Coulter representative for replacement product requests; (4) For customers in the U.S. and Canada, complete the attached Replacement Order Form; and (5) For customers outside the U.S. and Canada, contact their local Beckman Coulter representative. The consignee was requested to share the letter information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. If the affected product was forwarded to another laboratory, they were requested to provide them a copy of the letter. The consignee was requested to respond within 10 days either electronically via email or manually by completing and returning the enclosed Response Form.
Root Cause
Vendor change control
Distribution Pattern
Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Bahrain, Bangladesh, Bolivia, Brazil, Canada, China, Croatia, Czech Republic, Denmark, Egypt, El Salvador, Georgia, Greece, Hong Kong, India, Kazakhstan, Mexico, Namibia, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Taiwan, Thailand, Turkey, U.A.E, Vietnam, and Zimbabwe.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.