TriActiv ProGuard Embolic Protection System, Part number 61000-01. The product is shipped as a procedure kit. It is used with a standard 6F guide catheter using the standard femoral approach.
FDA Recall Z-1305-06 · Posted 2006-07-28
Recalling FirmKensey Nash Corp
Recall StatusTerminated
Product CodeNFA
Date Initiated2006-04-06
Date Terminated2006-08-10
Product Quantity69 kits
Reason for Recall
Alarm activiation-A priming issue involving the flow control unit due to a software problem has caused false postive Extraction Line Block (ELB) alarming. Use of this product may result in a health hazard resulting from early termination of the flush and extraction procedure after stenting
Firm Action
The recalling firm issued a draft recall letter to their European distributor who translated the letter for their customers.
Root Cause
Other
Distribution Pattern
The product was shipped to hospitals in Germany and Italy.