Penner Patient Stretcher/Transfer Lift System Models 391000-1 and 392000-1 The Pacific Stretcher/Transfer Lift is used with Whirlpool or Aqua-Aire (air bubbling) bathing system intended for use in nursing homes, hospitals, and assisted living facilities to transfer and/or lift patients under the direct supervision of trained staff. Model 392000-1 is equipped with a scale.

FDA Recall Z-1306-2015 · Posted 2015-03-20

Recalling FirmPenner Mfg Inc
Recall StatusTerminated
Product CodeMMP
Date Initiated2015-01-30
Date Terminated2016-03-10
Product Quantity158

Reason for Recall

If the patient sits on either wing, not squarely on the center section of the Stretcher Lift, tipping and injury to the operator or patient could result. The device manual was updated with an additional warning statement.

Firm Action

Beginning on 1/30/2015, the recalling firm sent letters via certified mail to their customers informing them of the change to the stretcher manual along with a revised manual which incorporated those changes. For questions regarding this recall call 402-694-5003.

Root Cause

Use error

Distribution Pattern

Worldwide Distribution - USA (nationwide) and Internationally to Canada.