MultiDiagnost Eleva Part Numbers: 9896 001 41501 and 9896 001 41502
FDA Recall Z-1307-04 · Posted 2004-08-04
Advertisement
Recall StatusTerminated
Product CodeIZI
Date Initiated2004-06-08
Date Terminated2006-02-09
Product Quantity8
Reason for Recall
When reviewing images in the module before sending them to PACS or printer the order can be erratic and out of sequence and may cause confusion with the interpretation of the examination.
Firm Action
The firm mailed letters via certified mail to their affected customers beginning on June 8, 2004. The letters instruct their customers to modify the default display on the ViewForum to exam/series date and time until a service representative can install the upgraded software R3.2.
Root Cause
Other
Distribution Pattern
Philips Medical Systems North American Company distributes product to medical facilities nationwide.