Coated VICRYL (Polyglactin 910) Suture, Ethicon Inc., A Johnson & Johnson Company, Somerville, NJ 08876 Sutures are intended for general soft tissue approximation and/or ligation.

FDA Recall Z-1309-2009 · Posted 2009-05-12

Recalling FirmEthicon, Inc.
Recall StatusTerminated
Product CodeGAM
Date Initiated2009-02-02
Date Terminated2009-10-21
Product Quantity1,758,360 total

Reason for Recall

Sterility Issue: The firm decided to remove the product because it cannot assure the sterility of the products due to a failure in the sterilization process.

Firm Action

Urgent Voluntary Product Recall notification letters were sent via UPS Next Day Air with receipt on February 2, 2009. Letters were sent to customers, Directors of Surgical Services, Director of Risk Management, Materials Manager and Distributors. The letters requested that customers immediately discontinue use of the products with lot numbers listed. Customers are to identify and return only the boxes identified in the attached tables. The affected product should be returned to Ethicon via the pre-paid authorized shipping label. Customers that are not returning affected product should mail the enclosed business reply card to Ethicon. Customers should direct questions regarding this recall to 1-888-202-3694 or their Ethicon sales representative. 5/13/09: It was discovered that a lot of PDS II sutures that had been previously identified as affected but not distributed, were mistakenly distributed. Revised letters including this number were sent to 25 customers on 5/13/09.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution -- USA, Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, Greece, Hong Kong, Israel, Japan, Korea, Luxemburg, Mexico, New Zealand, Norway, Panama, Puerto Rico, Sweden, Taiwan, United Kingdom, Uruguay, and Venezuela.