Terminated

Merit MAK Mini Access Kit, 4 French (1.3mm)

FDA Recall Z-1310-04 · Posted 2004-08-04

Recall StatusTerminated
Product CodeDRE
Date Initiated2004-07-20
Date Terminated2005-01-21
Product Quantity1924 kits

Reason for Recall

Firm received a report of an introducer sheath separating from its hub while in the patient. Surgery was successful in removal, even though the sheath kept breaking during the attempt.

Firm Action

Consignees were notified by phone and letter on 7/20/2004.

Root Cause

Other

Distribution Pattern

Nationwide, including VA facilities in MA, MI, NY, RI and TX. Argentina, Canada, China, Germany, Italy, Israel, Netherlands, New Zealand, Saudi Arabia, Scotland, Singapore, Thailand, United Arab Emirates.