TEG 5000 Functional Fibrinogen Reagent, Item No. 07-034

FDA Recall Z-1310-2022 · Posted None

Recall StatusTerminated
Product CodeJPA
Date Initiated2022-04-26
Date Terminated2024-03-15
Product Quantity3933 kits (US); 2429 kits (OUS)

Reason for Recall

Due to a shift in the citrated blood reference range for a normal population, the ranges stated in the IFU are changing. If the clinician relies solely on affected results in comparison to the normal donor reference range, it could lead to misdiagnosis and incorrect treatment or failure to treat.

Firm Action

On April 26, 2022, affected customers were provided with a recall letter via email. Customers may continue to use the TEG 5000 Functional Fibrinogen reagent with the current IFU in conjunction with the letter which reflects the updated ranges. Haemonetics will provide a revision to the IFU. Additionally, customers using Functional Fibrinogen reagent are being reminded to continue to follow the instructions in the current Instructions for Use (118474-MULTI (XX)), which state - Results from the TEG analyzer should not be the sole basis for a patient diagnosis TEG analyzer results should be considered along with a clinical assessment of the patient's condition and other coagulation laboratory tests.

Root Cause

Other

Distribution Pattern

Domestic distribution US Nationwide.