Terminated

Zoll Medical X Series, PropaqMD, PropaqM Defibrillator/Pacemaker/Monitor With System Software Version 02.10.02.00 or Higher, automated external defibrillator.

FDA Recall Z-1311-2014 · Posted 2014-03-31

Recall StatusTerminated
Product CodeMKJ
Date Initiated2014-03-05
Date Terminated2021-01-29
Product Quantity3,169 units

Reason for Recall

Defibrillator/Pacemaker/Monitor May Latch in a Continuous Device Reset Loop and cannot be used

Firm Action

Zoll Medical issued a Notification letter daetd March 5, 2014, to all affected consignees. The notification letter provides information on how to avoid the issue and instructs customers to contact ZOLL to get affected devices updated. Customers will be given the option to have a software upgrade kit sent to them to update their devices or return devices to ZOLL Service department in Chelmsford Massachusetts to have the device upgraded with the latest software release (Software version 02.16.04.00). Customers with questions were instructed to contact the 24/7 technical support numbers 1 (800) 348 9011 or +1 (978) 421 9460 to assist users with any aspect of this notice.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - USA (nationwide) and Internationally to Canada, AE, Afganistian, Australia, Belgium, Czech Republic , China, Chile, Croatia, Denmark, France, Great Britain, Hungary, Israel, Italy, Japan, Korea, Kuwait, Russia, Saudi Arabia, Singapore, South Africa,Thailand, and Turkey.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.