Fabius GS Anesthesia Machine; Gas machine for anesthesia or analgesia; Catalog Number: 8604700; Drager Medical Inc, Telford, PA 18969.
FDA Recall Z-1315-2008 · Posted 2008-07-16
Advertisement
Recalling FirmDraeger Medical, Inc.
Recall StatusTerminated
Product CodeBSZ
Date Initiated2008-02-05
Date Terminated2008-09-09
Product Quantity196 UNITS
Reason for Recall
Detachment of Component: The chassis on some machines were manufactured out of specifications. As a result, the caster may break loose from the chassis.
Firm Action
Consignees were notified by an Urgent Medical Device Recall letter sent on 2/7/08. The letter informed users that the recalling firm will contact them to schedule appointments to inspect any potentially affected machines. The letter instructed users to continue using the machines, before the inspection, making sure to carefully follow the warnings in the instruction manual. If a caster on the machine appears unstable before the inspection is conducted, users were advised to contact the recalling firm. For additional information, contact 1-800-543-5047.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Nationwide Distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.