Terminated

The LCS COMPLETE P.F.C.TM SIGMA RP Mobile Bearing Total Knee System - Product Usage: consists of three components intended to replace the femoral, tibial, and patellar articular surfaces of the knee joint.

FDA Recall Z-1315-2021 · Posted None

Recall StatusTerminated
Product CodeNJL
Date Initiated2021-02-22
Date Terminated2023-10-11
Product Quantity102 units total (Domestic: 0 units, Foreign: 162 units)

Reason for Recall

Globally the company is removing 37 lots affected lots for MBT & ATTUNE (S+) Tibial Bases due to potential damaged packaging. These implant components are utilized in Total Knee Arthroplasty. The affected tibial bases are designed to be used in cemented applications.

Firm Action

On 02/25/2021, Depuy Synthes issued an Urgent Medical Device Recall notices to customers via email. DePuy (Ireland) UC is initiating a voluntary removal for the 12 lots for ATTUNE (S+) Tibial Bases due to potential damaged packaging. These implant components are utilized in Total Knee Arthroplasty. The affected tibial bases are designed to be used in cemented applications.

Root Cause

Packaging change control

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV and the countries of Brazil, Austria, Belgium, Denmark, Estonia, France, Germany, Great Britain, Ireland, Israel, Italy, Macedonia, Mauritius, Netherlands, Norway, Poland, Romania, Slovakia, Slovenia, South Africa, Sweden, Switzerland, United Arab Emirates, China, Hong Kong, Japan, New Zealand, Korea, Singapore, Thailand.

Official Source

View FDA Recall →