Medtronic DLP Vessel Cannula, Model Number REF 30000

FDA Recall Z-1316-2024 · Posted 2024-03-19

Recall StatusOpen, Classified
Product CodeDWF
Date Initiated2024-02-07
Product Quantity4343 units

Reason for Recall

Medtronic is writing to inform you of incorrect labeling for three manufactured lots of the DLP Vessel Cannulae for the model and lot numbers listed above.

Firm Action

Medtronic Cardiac Surgery issued an Urgent: Medical Device Recall to its consignees on 02/07/2024 via UPS. The notice explained the issue, potential risk, and requested consignees identify and quarantine affected product, and arrange to return the affected product by contacting customer service at 1-800-854-3570, option 1, then Option 4. The notification should be shared with others in your organization as appropriate. If product was forwarded to another facility, please notify the facility of this Medtronic Urgent Medical Device Recall.

Root Cause

Packaging change control

Distribution Pattern

US and Brazil, Canada, Republic of Georgia, Germany, India, Ireland, Italy, Malaysia, South Africa, UK