SMR TT Hybrid glenoid Std - Peg S, REF: 1379.59.210
FDA Recall Z-1317-2023 · Posted 2023-04-01
Advertisement
Recalling FirmLimacorporate S.p.A
Recall StatusOpen, Classified
Product CodeMBF
Date Initiated2023-02-21
Product Quantity1
Reason for Recall
Shoulder glenoid fossa prosthesis was manufactured without tantalum markers so in the absence of the markers, on already implanted devices, a correct radiographic assessment of the hybrid glenoid position can be performed by the evaluation of the joint space, central TT peg position and cement mantel observation.
Firm Action
On 2/21/23, a recall notice was distributed to a customer who was asked to do the following:
1) Check your stock to locate and quarantine the affected devices. Devices must be
sent back.
2) Complete, and return the response form to [email protected]
3) The recall notice should be shared with anyone who needs to be aware within your
organization or shared with any organization where the potentially affected products
have been transferred.
Inquiries can be sent to [email protected]
Root Cause
Process control
Distribution Pattern
US: MO
Related FDA 510(k) Records