Access Immunoassay Systems Accu Tnl Reagent Kit when used in conjunction with UniCel Dxl Access Immunoassay Systems; Reagent Part Numbers: A78803 & 33340; Calibrator Part Number: 33345 The intended use: The Access AccuTnI assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma using the Access Immunoassay Systems to aid in the diagnosis and treatment of myocardial infarction and cardiac muscle damage. Cardiac Troponin I determination also aids in the risk stratification of patients with unstable angina or non-ST segment elevation acute coronary syndromes with respect to relative risk of mortality, myocardial infarction, or increased probability of ischemic events requiring urgent revascularization procedures.
FDA Recall Z-1318-2010 · Posted 2010-04-12
Advertisement
Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeMMI
Date Initiated2010-02-06
Date Terminated2012-05-09
Product Quantity196,071 kits
Reason for Recall
The recall was initiated because Beckman Coulter has confirmed customer reports that different results have been obtained using
the same patient samples on Access/Access 2 and UniCel Dxl platforms. Values obtained with
UniCel Dxl systems have been demonstrated to have a positive bias compared to values
obtained with Access or Access 2 systems.
Firm Action
An "URGENT: PRODUCT CORRECTIVE ACTION (PCA)" letter dated February 5, 2010, was sent to the customers. The letter describes the product, problem and action to be taken by the customers. The customers are recommended to immediately discontinue using AccuTnI Reagents and Calibrators on the UniCel DxI system until further notice. If this is not possible, the customers are directed to notify their stakeholders of the bias until further notice. The customers should complete, fax (786-639-4000) or mail the enclosed RESPONSE FORM.
If consignees need assistance or have any questions regarding the notification, they were instructed to contact Technical
Support at 1-800-854-3633 in the United States and Canada.
Root Cause
Other
Distribution Pattern
Worldwide distribution: USA and Canada