Terminated

Refractor Suspension Arm (P/N: 525004, 05 and 525006, 07) The Intended use of this medical device, as part of system to store and position ophthalmological examination and diagnostic equipment, is to store and position for examination a device commonly known as a refractor.

FDA Recall Z-1318-2018 · Posted None

Recall StatusTerminated
Product CodeHKN
Date Initiated2017-12-06
Date Terminated2019-05-02
Product Quantity933 ( of which 97 has been quarantined)

Reason for Recall

Product has been found with a missing Washer, Retaining Ring, and/or Stop Screw. Recalling firm was notified by distributor that Refractor Suspension Arm fell on a patient s lower torso along with the arm housing

Firm Action

Recalling firm sent letter to customers via certified mail for US customers and via regular mail to foreign customers. Verify product has the presence of the one or more of the three parts: Refractor Suspension Arm missing Screw, Washer and/or Retaining Ring (Snap Ring) that may be missing, and to assure it is functioning properly. For further questions, please call (513) 398-3937

Root Cause

Employee error

Distribution Pattern

Worldwide Distribution - USA(nationwide) Distribution to the states of : NJ, CA, OH, VA, MA, MN, IL , UT, TN, AR, LA, AZ, PA, TX, IA and HI., and to the countries of : Costa Rica, Jamaica, South Africa, New Zealand, South Korea, Dubai, Saudi Arabia and Jordan.