D711, Reference Membranes, Part #942-058, used with the ABL835 Blood Gas Analyzers.
FDA Recall Z-1319-06 · Posted 2006-08-02
Advertisement
Recalling FirmRadiometer America Inc
Recall StatusTerminated
Product CodeCHL
Date Initiated2005-06-17
Date Terminated2011-09-14
Product Quantity13 boxes/4 membranes per box
Reason for Recall
The recalled membranes cause the status value of the pH and electrolyte electrodes to be offset causing readings outside of the acceptable range.
Firm Action
The recalling firm notified their distributor/customers of this recall, via telephone on 6/17/2005, at which time, the customers were asked to destroy any and all suspect reference membranes found on-site and at their subaccount customer locations.
Root Cause
Other
Distribution Pattern
Nationwide, including TX, OK, MI, CO, NY, and Canada.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.