ThermoFisher Scientific Microm STP 420D automated tissue processor.

FDA Recall Z-1320-2010 · Posted 2010-04-12

Recall StatusTerminated
Product CodeIEO
Date Initiated2009-06-03
Date Terminated2011-06-29
Product Quantity16

Reason for Recall

The firm has received complaints regarding specimen integrity (higher rate of specimen fragmentation) as well as complaints that the chamber may stop rotating prematurely, resulting in sample deterioration.

Firm Action

Consignees were notified of the problem by letter, on or about 6/3/09, updating the instructions for use. The units were subsequently upgraded on or before 9/8/09. For additional questions and information, please contact ThermoFisher's Technical Support department , Subject: PT Module Advisory Notice.

Root Cause

Other

Distribution Pattern

Hawaii, Kentucky, Louisiana, Michigan, Missouri, Nebraska, New York, North Carolina, Ohio, Pennsylvania, Tennessee, Texas and West Virginia.