Terminated

TOPAZ MicroDebrider 0.8 mm REF AC4040-01; TOPAZ MicroDebrider IFS 0.8 mm REF ACH4040-01; TOPAZ MicroDebrider 0.8 mm REF Q6000-01; TOPAZ EPF MicroDebrider 0.8 mm 45 REF Q6002-01

FDA Recall Z-1321-2016 · Posted 2016-04-05

Recall StatusTerminated
Product CodeGEI
Date Initiated2016-03-03
Date Terminated2017-05-17
Product Quantity26,903 devices

Reason for Recall

During functional testing of devices after real time aging, it was determined that product failed to meet shelf-life at the expiration date indicated on the product labels.

Firm Action

Consignees were notified via letter on 3/3/2016. Consignees were instructed to quarantine the recalled products and complete an Inventory Return Certification Form.

Root Cause

Package design/selection

Distribution Pattern

Worldwide Distribution-US (nationwide) and the following countries-Austria, Belgium, Bulgaria, Canada, Denmark, Estonia, Finland, France, Germany, Gibraltar, Great Britain, Greece, Ireland, Israel, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, China, Hong Kong, India, Lebanon, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arabic Emirates, Vietnam, Australia, Japan, New Zealand, Brazil, and Chile.