Reference Electrodes for ABL 700/800 Blood Gas Analyzers. Electrode Part Number: 945-603, distributed one (1) electrode per box.

FDA Recall Z-1323-06 · Posted 2006-08-03

Recall StatusTerminated
Product CodeCHL
Date Initiated2005-09-19
Date Terminated2012-02-24
Product Quantity34

Reason for Recall

Reference electrodes for the firm's ABL 700/800 Blood Gas Analyzers are oversized and will not fit into the instrument.

Firm Action

Following the receipt of three (3) customer complaints, on 9/14/2005, the firm issued an E-Mail notification [Management Cover] to their Service Personnel on 9/19/2005. The Management Cover instructed the service personnel to check the electrodes located at their stock and their customer locations; put all electrodes, which cannot be inserted in their call-back stock; and provide for credit to their customers through normal channels. The Service Personnel in turn, contacted their customers via telephone on or around 9/20/2005 and asked that they check their inventories for the presence of any electrodes which would not fit in the analyzer and return any such electrodes to Radiometer America for replacement

Root Cause

Other

Distribution Pattern

Nationwide and Canada.