Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Medtronic Grafton Crunch a) DBM S44105 GRAFTON 5CC CRUNCH, REF S44105; b) DBM S44115 GRAFTON 15CC CRUNCH, REF S44115; c) DBM T44105 5CC GRAFTON CRUNCH, REF T44105; d) DBM T44105INT 5CC GRAFTON CRUNCH, REF T44105INT; e) DBM T44115 15CC GRAFTON CRUNCH, REF T44115; f) DBM T44115INT 15CC GRAFTON CRUNCH, REF T44115INT; Demineralized Bone Matrix (DBM) product

FDA Recall Z-1324-2023 · Posted 2023-04-04

Recall StatusOpen, Classified
Product CodeMBP
Date Initiated2023-02-08
Product Quantity5308 units

Reason for Recall

Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.

Firm Action

Medtronic Sofamor Danek notified customers of the recall on about 02/08/2023, via letter. Customers were instructed to identify and quarantine any affected products, return any affected products, complete and return the Customer Confirmation Form, and provide the letter to all personnel that should be aware or where affected products have been transferred/distributed.

Root Cause

Process control

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Brunei Darussalam, Canada, Chile, Colombia, Denmark, Dominican Republic, Ecuador, Finland, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Israel, Jamaica, Japan, Jordan, Korea, Republic Of, Kuwait, Lebanon, Liechtenstein, Luxembourg, Macao, Malaysia, Mexico, Mongolia, Namibia, Netherlands, New Zealand, Norway, Oman, Panama, Paraguay, Peru, Philippines, Portugal, Puerto Rico, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, Trinidad And Tobago, United Arab Emirates, United Kingdom, Viet Nam.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑