Terminated

Fresenius 2008 Series Hemodialysis Delivery Systems: 2008T, 2008K2, 2008K and 2008K@home: Models: Fresenius 2008K and 2008K2 Dialysate Delivery System Fresenius 2008K@home Hemodialysis Machine with bibag System 2008T Hemodialysis Machine The Fresenius 2008 Series Machines are indicated for acute and chronic dialysis therapy

FDA Recall Z-1325-2016 · Posted 2016-04-05

Recall StatusTerminated
Product CodeKDI
Date Initiated2016-03-10
Date Terminated2017-03-09
Product Quantity8384 units (8369 US) (15 OUS)

Reason for Recall

Added Warning Statement: Fresenius 2008 Series Hemodialysis Machines False Blood Leak Alarm When Dialyzing Patients Treated with Hydroxocobalamin (or any form of Vitamin B-12)

Firm Action

Fresenius Medical Care issued an Urgent Medical Device Alert dated March 10, 2016, to all affected consignees. Consignees were instructed to place notification and Warning Addendum with their 2008 Series Hemodialysis Machine Operators Manual. A reply form was included as part of the customer communication with instructions to fill out and return via fax or scanned/emailed copy. Further information or support concerning this issue, customers were instructed to contact Fresenius Technical Services at 800-227-2572.

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution