Terminated

Abutment Screw Retrieval Instrument RP/WP/6.0 - Product Usage: The instruments for abutment screw retrieval can be used if an abutment screw is broken and a part of the screw is left inside the implant screw channel. If the implant thread needs to be cleaned before placing a new screw, a screw tap repair can be used. The instruments are intended for both simple and advanced treatments.

FDA Recall Z-1325-2020 · Posted 2020-02-19

Recalling FirmNobel Biocare Usa Llc
Recall StatusTerminated
Product CodeNDP
Date Initiated2019-03-14
Date Terminated2024-02-15
Product Quantity132 units

Reason for Recall

Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.

Firm Action

On 3/27/2018 the firm issued a recall notice by email to affected consignees. The recall notification informed the following actions to be taken by the consignees: 1. Identify affected devices as detailed in the Customer Reply Form. 2. Do not use affected devices. 3. Send back the Customer Reply Form and the affected devices. The firm will replace affected devices free of charge. This letter needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. We kindly ask you to follow instructions provided in this Recall Letter and return the attached Customer Reply Form to the firm via email or fax at 714- 282-5073.

Root Cause

Packaging process control

Distribution Pattern

Worldwide distribution - US Nationwide distribution and countries of CA, AU, JP, BR, ZA, RU, TW, HK, IN, CN, SA, NZ, MO, CO, LB, AT, BE, CH, CY, DE, DK, EE, ES, FI, FR, GB, GG, GR, HR, HU, IE, LT, LU, NL, PL, PT, RS, SE.

Official Source

View FDA Recall →