Recall StatusTerminated
Date Initiated2019-09-26
Date Terminated2020-12-03
Product Quantity76 units
Reason for Recall
Power supply (Model Number: TLC-2003) ME90A2403F01 (24 VDC, 3.75 A, 90 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 4th edition specification) shipped versus Power supply (Model Number: TLC-2003) MENB1100A2403F01 (24 VDC, 4.2 A, 101 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 3rd edition specification)
Firm Action
Email was sent out to healthcare practitioners regarding the recall on September 26, 2019.
Healthcare practitioners are instructed to courier their TLC-2000 to Theralase. A Field Inspection Test will be completed and once passed, a Special Inspections for Medical Electrical Equipment and Systems label will be applied to the TLC-2000.The Field Inspection Test will be completed according to either CSA 60601-1 or UL 60601-1 dependent upon the customers Canadian or United States location, respectively.
Root Cause
Process control
Distribution Pattern
Worldwide distribution - US Nationwide and countries Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania.
Official Source