Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Fluoroscan Mini C-arm InSight2, software versions 3.x, 4.0, 4.0.1, 4.0.2 The device is a Mini C-arm Fluroscopic Imaging System designed to provide physicians with general fluoroscopic visualization of a patient, including but no limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation.

FDA Recall Z-1327-2013 · Posted 2013-05-17

Recalling FirmHologic, Inc.
Recall StatusTerminated
Product CodeJAA
Date Initiated2013-04-25
Date Terminated2014-08-21
Product Quantity2780

Reason for Recall

The audible alarm to alert the operator when cumulative irradiation time reaches five minutes is not accurate and the total cumulative irradiation time is not accurate.

Firm Action

The firm, Hologic, Inc. sent an "Urgent: Medical Device Recall" letter dated April 25, 2013, via FedEx to consignees/customers notifying them of the software defect and providing them with corrected user installable software version. The customers were instructed to complete and return the Enclosure to Customer Notification, Confirmation of Notification form via fax to Hologic at 866-652-8674 or mail the document to Hologic using the enclosed pre-addressed, pre-paid return envelope. Should you have any questions regarding this recall, require further information or assistance with the software installation call Hologic's Help Desk at 1-800-321-4659.

Root Cause

Software design

Distribution Pattern

Worldwide distribution: USA (nationwide) and countries of: DE, RU, AU, IT, GB, ES, FR, KR, ZA, CA, AR, LT, TW, CN, BE, AT, JP, QA, SA, SG, HK, SI, NL, and SE.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑