Open, Classified

Medtronic Grafton Matrix: a) DBM S42200 GRAFTON 2.5CMX10CM 2EA MATRIX, REF S42200; b) DBM S42210 GRAFTON 2.5CMX5CM 2EA MATRIX, REF S42210; c) DBM T42200 2.5CMX10CM 2 EACH GRAFTON MAT, REF T42200; d) DBM T42200AUS 2.5CMX10CM 2 EA GRAFTON MX, REF T42200AUS; e) DBM T42200INT 2.5CMX10CM 2 EA GRAFTON M, REF T42200INT; f) DBM T42200JPN GRAFTON MATRIX 2.5CM X10CM, REF T42200JPN; g) DBM T42210 2.5CMX5CM 2 EACH GRAFTON MATR, REF T42210; h) DBM T42210AUS 2.5CMX5CM 2 EA GRAFTON MX, REF T42210AUS; i) DBM T42210INT 2.5CMX5CM 2 EA GRAFTON M, REF T42210INT; j) DBM T42210JPN GRAFTON MATRIX 2.5CM X 5CM, REF T42210JPN; Demineralized Bone Matrix (DBM) product

FDA Recall Z-1327-2023 · Posted 2023-04-04

Recall StatusOpen, Classified
Product CodeMBP
Date Initiated2023-02-08
Product Quantity19057 units

Reason for Recall

Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.

Firm Action

Medtronic Sofamor Danek notified customers of the recall on about 02/08/2023, via letter. Customers were instructed to identify and quarantine any affected products, return any affected products, complete and return the Customer Confirmation Form, and provide the letter to all personnel that should be aware or where affected products have been transferred/distributed.

Root Cause

Process control

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Brunei Darussalam, Canada, Chile, Colombia, Denmark, Dominican Republic, Ecuador, Finland, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Israel, Jamaica, Japan, Jordan, Korea, Republic Of, Kuwait, Lebanon, Liechtenstein, Luxembourg, Macao, Malaysia, Mexico, Mongolia, Namibia, Netherlands, New Zealand, Norway, Oman, Panama, Paraguay, Peru, Philippines, Portugal, Puerto Rico, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, Trinidad And Tobago, United Arab Emirates, United Kingdom, Viet Nam.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑