Pointe Scientific Liquid Glucose (HEXO) Reagent Set, for the quantitative determination of glucose in serum, 10 x 100 ml R1, 10 x 20 ml R2 glucose reagent, or 1 x 2 L R1 x 500 ml R2, for use with Hitachi analyzer; in vitro diagnostic, Catalog numbers HG920-1200 and HG920-2500.

FDA Recall Z-1329-06 · Posted 2006-08-03

Recall StatusTerminated
Product CodeCFR
Date Initiated2006-06-07
Date Terminated2007-11-19
Product Quantity153 kits

Reason for Recall

Product does not meet performance specifications through its labeled expiration period.

Firm Action

Customers were notified via recall letter dated 6/7/06 to cease using and to discard these lots, and to send back the recall response form for replacement stocks.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, Canada, China, Panama, Russia and Trinidad.