CombiDiagnost PCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.

FDA Recall Z-1329-2020 · Posted 2020-02-20

Recall StatusTerminated
Product CodeJAA
Date Initiated2019-12-19
Date Terminated2020-06-18
Product Quantity131

Reason for Recall

When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.

Firm Action

On December 19, 2019, the firm initiated mailing of the Medical Device Recall Letter to all affected customers. Customers were advised to use the system according to the Instructions for Use (IFU). In particular: Error 80 issue: The issue can be prevented if the operator avoids tilting the table by Table Up/Down joystick and instead uses the Table Tilt Left/Right joystick. If the issue occurs, the operator should remove the patient from the table and bring him/her to upright position so that the contrast agent does not enter the brain. SPDU issue: The user should pay attention to whether it smells of smoke in the room. If this is the case, immediately engage the system emergency stop button on the table or geometry control console, remove patient, stop using the system for further examinations and contact customer service. Philips plans to install a system software version upgrade and to perform an on-site check of the SPDU thermos contact at affected systems, which will eliminate these issues. A Philips Service Engineer will contact you when the Field Action Kit is available to be implemented. Should you need to communicate with Philips with regard to this program, please reference Field Change Order (FCO) 70900045 if you have a CombiDiagnost GCF system and FCO 70900048 if you have a CombiDiagnost PCF system. Technical Support Line: 1-800-722-9377.

Root Cause

Software design

Distribution Pattern

Worldwide distribution US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MI, MN, MO, MT, NC, NJ, NY, OH, TX, UT, VA and countries of Canada, Australia, Austria, Bangladesh, Belgium, Bulgaria, Chile, China, Colombia, C¿te D'Ivoire, Cyprus, Czech Republic, Dominican Republic, Egypt, Ethiopia, France, Germany, Ghana, Hungary, Iran, Italy, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Martinique, Mexico, Nepal, Netherlands, New Zealand, Nigeria, Oman, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Serbia, South Africa, Spain, Switzerland, Tanzania, Thailand, Turkey, United Arab Emirates, United Kingdom, Viet Nam.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.