Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Medtronic Accelerate Grafton DBF a) ACCELERATE BG SET T50203 GRAFTON DBF 3CC, REF T50203; b) ACCELERATE BG SET T50206 GRAFTON DBF 6CC, REF T50206; c) ACCELERATE BG SET T50209 GRAFTON DBF 9CC, REF T50209; d) ACCELERATE BG SET T50212 GRAFTON DBF12CC, REF T50212; Demineralized Bone Matrix (DBM) product

FDA Recall Z-1332-2023 · Posted 2023-04-04

Recall StatusOpen, Classified
Product CodeMBP
Date Initiated2023-02-08
Product Quantity3037 units

Reason for Recall

Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.

Firm Action

Medtronic Sofamor Danek notified customers of the recall on about 02/08/2023, via letter. Customers were instructed to identify and quarantine any affected products, return any affected products, complete and return the Customer Confirmation Form, and provide the letter to all personnel that should be aware or where affected products have been transferred/distributed.

Root Cause

Process control

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Brunei Darussalam, Canada, Chile, Colombia, Denmark, Dominican Republic, Ecuador, Finland, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Israel, Jamaica, Japan, Jordan, Korea, Republic Of, Kuwait, Lebanon, Liechtenstein, Luxembourg, Macao, Malaysia, Mexico, Mongolia, Namibia, Netherlands, New Zealand, Norway, Oman, Panama, Paraguay, Peru, Philippines, Portugal, Puerto Rico, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, Trinidad And Tobago, United Arab Emirates, United Kingdom, Viet Nam.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑