Directional Laser Probe with Alcon / Lumenis connector (27 gauge / 0.4 mm)
FDA Recall Z-1333-2024 · Posted 2024-03-21
Advertisement
Recalling FirmD.O.R.C. Dutch Opthalmic Research Center Intl B.V.
Recall StatusOpen, Classified
Product CodeHQC
Date Initiated2024-01-15
Product Quantity13 boxes x 6 units per box = 78 units
Reason for Recall
When using those products you may experience difficulties to extend or retract the laser fiber, and in some occasions may also have difficulties to direct the
laser fiber tip precisely.
Firm Action
An Urgent Medical Device Recall notification letter was sent to customers on 01/15/2024. Actions to be Taken by the Customer: Pass this letter on to all those who need to be aware within your organization and/or to any organization where the potentially affected devices have been transferred to. Verify whether you have any unopened boxes, or individual pouches of the affected DORC directional laser probes in your inventory. Remove any remaining boxes and individual Products from your inventory and return the impacted Product to DORC following the instructions in the attachment. Complete, even if you do not return the Product, the on-line reply form per instructions in the attachment. In case of questions please contact us in any of the following manners: Contact your local DORC representative, or call the Customer Service at our distributor Dutch Ophthalmic, USA at 1-800-753-8824 x 1, or send an email to [email protected].
Root Cause
Process control
Distribution Pattern
US Nationwide distribution in the states of AZ, CO, FL, IN, MD, NC, NY, OH, SC, TX and PR.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement