ABX Pentra 120,120R,120DX, automated hematology analyzer
FDA Recall Z-1334-06 · Posted 2006-08-03
Advertisement
Recalling FirmABX Diagnostics Inc
Recall StatusTerminated
Product CodeGKZ
Date Initiated2006-03-24
Date Terminated2012-05-23
Reason for Recall
Labels can be incorrectly placed on a sample tube resulting in a possible error in results attributed to a sample.
Firm Action
Notifications have been prepared advising that following an occurrence observed outside of the United States, HORIBAABX is providing information regarding the rare circumstance of a label being incorrectly placed on a sample tube, resulting in a possible error in results attributed to a sample. The March 24, 2006, letter identifies various conditions in which the anomaly occurs.
In order to prevent this type of incident, the firm recommends that the customer ensure that the identification label is firmly attached to the sample tubes during their positioning or removal from the racks.
Root Cause
Other
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.