Playtex Nurser Deluxe Double Electric Breast Pump, Model X06578AO, powered by a 12V DC power supply, Model SY-12120, which is included with the package.

FDA Recall Z-1336-2014 · Posted 2014-04-04

Recall StatusTerminated
Product CodeHGX
Date Initiated2014-03-18
Date Terminated2016-06-10
Product Quantity35,803

Reason for Recall

Some of the power adapters outer casings may become loose and separate, resulting in a potential for electric shock.

Firm Action

Energizer sent and Voluntary Retail Recall Notice letter dated March 18, 2014 to retail customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to stop sale and return the affected product. The letter provides instruction to customers to return the affected product. For questions or concerns, contact Energizer Personal Care at 1-888-207-1492 or visit www.playtexproducts.com. On 9/9/2014 the firm expanded the recall to one additional production lot to include: Suspect date codes for Breast Pumps: P12315-XXXX to P13205-XXXX. Suspect adaptor date codes: 1238 to 1324 Revised Press issued 9/9/2014.

Root Cause

Component change control

Distribution Pattern

USA Nationwide Distribution in the states of: AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, MD, MI, MN, MO, NC, NJ, NV, NY, OH, ON, OR, PA, SC, TN, TX, VA, VT, WA, WI Canada - one consignee, Energizer (recalling firm's) company store