Medtronic HVAD Controller, Product Numbers: 1400, 1401, 1401AU, 1401BR, 1401CA, 1401CH, 1401DE, 1401GB, 1401IL, 1401IN, 1401IT, 1401JP, 1401US, 1403US, 1407AR, 1407AU, 1407BR, 1407CA, 1407CA-CLIN, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR

FDA Recall Z-1336-2021 · Posted None

Recalling FirmHeartware, Inc.
Recall StatusOpen, Classified
Product CodeDSQ
Date Initiated2021-02-26
Product Quantity56,018 units (28,449 US)

Reason for Recall

There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.

Firm Action

The firm issued a Medical Device Notice to its consignees on 02/26/2021 by letter delivered via UPS 2-day delivery. The letter explained the issue and directed the physician to review the patient management recommendations on the frequency of of inspecting the HVAD Controller, HVAD Controller power cables, and HVAD Monitor data cables. For questions regarding this material, please contact your Medtronic Field Representative.

Root Cause

Device Design

Distribution Pattern

Worldwide