ARCHITECT c8000 Processing Module; List Number 01G06-01. Manufactured by Toshiba Corporation, Medical Systems Division, Japan. Distributed by Abbott Laboratories Diagnostic Division, Irving, Texas

FDA Recall Z-1341-04 · Posted 2004-08-13

Recall StatusTerminated
Product CodeCGZ
Date Initiated2004-06-10
Date Terminated2004-11-19
Product Quantity171 units and 25 spare parts distributed

Reason for Recall

Process change in the manufacturing of the AC/DC driver board used on the analyzer may result in one or more of the solid state relays to fail during use.

Firm Action

The firm initiated the recall by Technical Service Bulletin sent to affected customers. The AC/DC boards will be replaced on each analyzer by a firm representative.

Root Cause

Other

Distribution Pattern

Product was distrbued nationwide and to the following foreign countries: Mexico, Australia, Germany, Hong Kong, Canada, Thailand, Grand Cayman, Chile, New Zealand, Columbia, Brazil, and Singapore.