UST-2265-2 Probe, Part Number: UST-2265-2 - Product Usage: The UST-2265-2 probe is intended to be used for diagnostic ultrasound imaging or fluid flow analysis of the human body as Continuous-wave Doppler (CWD) for Cardiac Adult, Cardiac Pediatric, and Peripheral Vessel.
FDA Recall Z-1341-2020 · Posted None
Advertisement
Recalling FirmHitachi Medical Systems America Inc
Recall StatusTerminated
Product CodeITX
Date Initiated2020-01-31
Date Terminated2020-07-31
Product Quantity18
Reason for Recall
These 18 probes may have mis-wiring in the inner cables, causing decreased sensitivity and thus resulting in lower peak velocity readings.
Firm Action
On January 31, 2020, the firm initiated customer communications via Urgent Medical Device Recall letters. Customers were informed of the potential mis-wiring issue. Clinicians are advised to discontinue use of the probes to minimize adverse events. Hitachi Service will contact each affected site to schedule a service visit. Hitachi will test all affected probes and remove and replace any probes that do not pass the test. The correction is expected to begin February 2020.
Root Cause
Process change control
Distribution Pattern
Domestic distribution to CA, FL, GA, MA, NV, OH, TX, and WV. Foreign distribution to Canada.