COHERENCE Oncologist, Model Nos. 07333680 & 07351898

FDA Recall Z-1346-2020 · Posted None

Recall StatusTerminated
Product CodeIYE
Date Initiated2020-01-16
Date Terminated2021-01-04
Product Quantity64 Worldwide/26 Domestic

Reason for Recall

A potential safety issue arises when an offset calculation is followed by a filter operation; correcting the image alignment after this sequence will result in incorrect offset values which could lead to incorrect repositioning of the patient and dose to wrong location.

Firm Action

The firm started notifying customers on January 16, 2020, with letters titled Urgent Medical Device Correction. Customers were advised of the product safety issue. To avoid the possibility of an incorrect alignment of the patient, the user must ensure that filtered images arrive on the COHERENCE Oncologist, or apply any additional filters to the image BEFORE using the "interactive shift" or "landmark registration" tools. The firm will be offering a software solution for the issue free of charge. The firm's letter stated that based on their investigation, customers may continue to use their devices.

Root Cause

Software Design Change

Distribution Pattern

Domestic distribution to CA, FL, GA, IA, MA, MI, NE, NJ, NV, NY, OH, OK, PA, SD, TN, WV. Worldwide foreign distribution.