Enteral Infusion Pump

FDA Recall Z-1347-2021 · Posted None

Recall StatusCompleted
Product CodeLZH
Date Initiated2021-01-19
Product Quantity146,944,334 units

Reason for Recall

The potential for air appearing in the enteral feed pumping set tubing during set-up.

Firm Action

On 03/01/2021, Cardinal Health issued an Urgent Medical Device Correction notice to customers via letter to the potential of air entering the feed set tubing of enteral fusion pumps with the specified product codes

Root Cause

Process control

Distribution Pattern

Domestic: All 50 states and DC, Foreign: Argentina, Australia, Brazil, Canada, Chile, Colombia, Japan, Mexico, Netherlands, Panama, Peru, Puerto Rico, Singapore, Switzerland, Turkey