Da Vinci S 8mm Instrument Cannula with outlet for use with the Da Vinci Surgical System; Model IS2000, Part Numbers: 400254-01, 400255-01; Used in endoscopic surgery to establish a port of entry for Intuitive Surgical's EndoWrist Instruments. Product is distributed by Intuitive Surgical, Inc., Sunnyvale, CA

FDA Recall Z-1348-2008 · Posted 2008-08-06

Recall StatusTerminated
Product CodeNAY
Date Initiated2008-01-04
Date Terminated2011-03-31
Product Quantity39 units

Reason for Recall

(1) Incorrect dimension on Luer on smoke evacuation cannulae not allowing for secure attachment function. (2) Incorrect labeling. External labeling lot number differs from lot number etched on the cannula.

Firm Action

On 1/4/08, the firm initiated the recall and its notification was via Urgent Device Recall letters explaining the reason for the recall, requesting immediate discontinued use, segregate non-conforming product in a secure area, and a Customer Service Rep will make contact in order to arrange for the retrieval of the material.

Root Cause

Process control

Distribution Pattern

Nationwide