TLC-2000 Therapeutic Medical Laser System Power Pack, Model Number TLC-2002
FDA Recall Z-1348-2020 · Posted 2020-02-13
Recalling FirmTheralase Inc.
Recall StatusTerminated
Product CodeNHN
Date Initiated2019-09-26
Date Terminated2020-12-03
Product Quantity202 units
Reason for Recall
Power Pack (component) was shipped with an internal battery pack tested to UN-38.3, UL-2054 and IEC-62133 (2nd Edition) standards; however, not tested to IEC-60601-1-2 (3rd edition) standards
Firm Action
Email was sent out to healthcare practitioners regarding the recall on September 26, 2019.
Please courier your TLC-2000 to Theralase, at your cost, for modification. A Field Inspection Test will be completed and once passed, a Special Inspections for Medical Electrical Equipment and Systems label will be applied to your TLC-2000 Power Pack..
Root Cause
Process control
Distribution Pattern
U.S.A.; Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania