Terminated

Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw

FDA Recall Z-1350-2020 · Posted None

Recall StatusTerminated
Product CodeKWP
Date Initiated2019-12-31
Date Terminated2025-11-17
Product Quantity179 units

Reason for Recall

Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.

Firm Action

Ulrich notified initially notified customers, via phone call, on 12/20/2019. On about 12/31/2019, customers were also sent an official "URGENT SAFETY INFORMATION: MEDICAL DEVICE CORRECTIONS & REMOVALS" letter. Customers were informed that the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation. Instructions included cease use of the affected devices, return them to Ulrich, and complete and return the provided "Product Recall" form. Customers were also instructed to inform all users of the devices within each organization are informed. Questions or concerns can be directed to Mr. Louis Milos, Sr. Manager Business and Quality Systems, at 636-519-0268 ext. 105, [email protected], Monday through Friday 8:00am to 5:00pm CST.

Root Cause

Process control

Distribution Pattern

AZ, TX, GA, FL, AR, CA

Official Source

View FDA Recall →