Easy Read App associated with First Response Pregnancy Test Sticks - Intended use of converting analog pregnancy test results into text pregnant or not pregnant for display on consumer devices (e.g., smart phone, iPad)
FDA Recall Z-1350-2023 · Posted 2023-04-07
Advertisement
Recalling FirmChurch & Dwight Inc
Recall StatusOpen, Classified
Product CodeLCX
Date Initiated2023-01-25
Product Quantity780000 downloads
Reason for Recall
Church & Dwight First Response Pregnancy Kits was marketed with the Easy Read App without a new 510k. There is also a potential for Easy Read App to be unable to reads the test results (i.e., no results reported on the APP screen) or to misread the analog results shown on the pregnancy test stick and may lead to misinterpretation of false positive or false negative tests by the user.
Firm Action
Church & Dwight Company Easy Read App requested the developer to remove the App from the Apple and Google Play app stores on 1/25/23 and third-party vendor disabled the App on consumers devices. Consumers who try to use the downloaded App receive a message that the App is no longer available.
Root Cause
No Marketing Application
Distribution Pattern
US Nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement