Foot Controls Model: S-N1 and S-N2 Product Usage: Variable foot control for use with drive units operating mechanical instruments.
FDA Recall Z-1351-2018 · Posted None
Advertisement
Recalling FirmW & H DentalWerk Burmoos GMBH - Site 1
Recall StatusTerminated
Product CodeEBW
Date Initiated2017-12-21
Date Terminated2019-06-07
Reason for Recall
During continuous postmarket surveillance activities, we have noticed a potential malfunction of individual foot controls of the type S-N1/S-N2.
In case the error occurs, the foot control might not switch back to zero position when the pedal is released. Consequently, the motor keeps rotating. If this happens during automatic thread cutting or automatic screwing-in of implants, it may, in the worst case, lead to a serious injury of the patient.
Firm Action
On December 22, 2017 a field safety notice was issued to customers notifying them of the potential problem with the foot controls.
On February 15, 2018 a replacement foot control was sent to customers with instructions to unplug the original foot control, cut the cable once disconnected to an electrical source and properly dispose of. The replacement foot control is to be used instead. Questions or concerns can be directed to [email protected]
Root Cause
Process change control
Distribution Pattern
US Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.