TLC-2000 Therapeutic Medical Laser System
FDA Recall Z-1354-2020 · Posted 2020-02-13
Advertisement
Recalling FirmTheralase Inc.
Recall StatusTerminated
Product CodeNHN
Date Initiated2019-09-26
Date Terminated2020-06-04
Product Quantity92 units
Reason for Recall
905 nm laser diodes may be below rated power output.
Firm Action
Email was sent out to healthcare practitioners regarding the recall on September 26, 2019.
.Healthcare practitioners are instructed to courier their TLC-2000 to Theralase for analysis. Their 905 nm laser diodes will be analyzed and recalibrated, as required, to meet rated power output.
Root Cause
Device Design
Distribution Pattern
USA; Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.