Terminated

ACCU-CHEK Advantage Complete Diabetes Monitoring Kit, Cat. No. 860. Ref. no. 080086000 with meter model number 870; Ref. no. 03002531001 with meter model number 33304510; Ref. no. 03002531001 with meter model number 2138930; Ref. no. 03002531001 with meter model number 2138930.

FDA Recall Z-1355-06 · Posted 2006-08-05

Recall StatusTerminated
Product CodeCGA
Date Initiated2006-06-22
Date Terminated2008-04-07
Product Quantity55,189 meters (for all Recalls Z-1354-1371-06)

Reason for Recall

The meter gives a strip error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a strip error and the manual only instructs the user to retest with a new strip using a larger drop of blood.

Firm Action

A recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.

Root Cause

Other

Distribution Pattern

Worldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands.