Open, Classified

Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff, and Cuffless: Disposable Inner Cannula or Reusable Inner Cannula (REF, Product Description): 10CN10H, 10.0MM SHILEY CUFFED TRACH CAN; 10CN10R, 10.0MM ADT FLEX TRACH W TG CUFF; 10UN10A, 10.0MM UNCUFF TRACH TUBE X1; 10UN10H, 10.0MM SHILEY CUFFLESS TRACH C; 10UN10R, 10.0MM ADT FLEX TRACH CUFFLESS; 4CN65A, 6.5MM TRACH TUBE W TG CUFF X1; 4CN65H, 6.5MM SHILEY CUFFED TRACH CANN; 4CN65R, 6.5MM ADT FLEX TRACH W TG CUFF X1; 4UN65A, 6.5MM UNCUFF TRACH TUBE X1; 4UN65H, 6.5MM SHILEY CUFFLESS TRACH CA; 4UN65R, 6.5MM ADT FLEX TRACH CUFFLESSX1; 5CN70A, 7.0MM TRACH TUBE W TG CUFF X1; 5CN70H, 7.0MM SHILEY CUFFED TRACH CANN; 5CN70R, 7.0MM ADT FLEX TRACH W TG CUFF X1; 5UN70A, 7.0MM UNCUFF TRACH TUBE X1; 5UN70H, 7.0MM SHILEY CUFFLESS TRACH CA; 5UN70R, 7.0MM ADT FLEX TRACH CUFFLESSX1; 6CN75A, 7.5MM TRACH TUBE W TG CUFF X1; 6CN75H, 7.5MM SHILEY CUFFED TRACH CANN; 6CN75R, 7.5MM ADT FLEX TRACH W TG CUFF X1; 6UN75H, 7.5MM SHILEY CUFFLESS TRACH CA; 6UN75R, 7.5MM ADT FLEX TRACH CUFFLESSX1; 7CN80A, 8.0MM TRACH TUBE W TG CUFF X1; 7CN80H, 8.0MM SHILEY CUFFED TRACH CANN; 7CN80R, 8.0MM ADT FLEX TRACH W TG CUFF X1; 7UN80A, 8.0MM UNCUFF TRACH TUBE X1; 7UN80H, 8.0MM SHILEY CUFFLESS TRACH CA; 7UN80R, 8.0MM ADT FLEX TRACH CUFFLESSX1; 8CN85A, 8.5MM TRACH TUBE W TG CUFF X1; 8CN85H, 8.5MM SHILEY CUFFED TRACH CANN; 8CN85R, 8.5MM ADT FLEX TRACH W TG CUFF X1; 8UN85A, 8.5MM UNCUFF TRACH TUBE X1; 8UN85H, 8.5MM SHILEY CUFFLESS TRACH CA; 8UN85R, 8.5MM ADT FLEX TRACH CUFFLESSX1; 9CN90H, 9.0MM SHILEY CUFFED TRACH CANN; 9CN90R, 9.0MM ADT FLEX TRACH W TG CUFF X1; 9UN90H, 9.0MM SHILEY CUFFLESS TRACH CA; 9UN90R, 9.0MM ADT FLEX TRACH CUFFLESSX1

FDA Recall Z-1357-2023 · Posted 2023-04-19

Recalling FirmCovidien
Recall StatusOpen, Classified
Product CodeJOH
Date Initiated2023-03-24
Product Quantity599,187

Reason for Recall

A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding.

Firm Action

On 3/24/23, recall notices were mailed to risk managers and directors of respiratory care. Tracheostomy Tubes that do not securely attach or stay attached to caps or accessories should not be used. 1) All affected unused product should be quarantined and returned to the recalling firm. 2) Complete and then return the customer confirmation form to [email protected] 3) Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected product from the specified lots has been transferred or distributed. 4) If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. Customers with questions can contact Quality Assurance at 800-255-6774, option 4.

Root Cause

Component change control

Distribution Pattern

Worldwide - US Nationwide distribution including in the states of PR, CA, FL, MN, PA, AZ, IL, TN, OH, MI, GA, KS, HI, MD, AK, NY, NE, IA, OK, VA, ND, MO, NC, SC, KY, OR, IN, WA, TX, WI, CT, MA, WV, CO, VT, NJ, SD, AR, AL, MS, LA, DE, MT, WY, DC, UT, NV, NH, RI, ID, NM, ME and the countries of Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bolivia, Plurinational State Of Bosnia And Herzegovina, Brazil, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Cyprus, Czech Republic, Denmark, El Salvador, Estonia, Finland, France, French Guiana, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, Indonesia, Ireland, Israel, Italy, Jordan, Kosovo, Kuwait, Lebanon, Lithuania, Luxembourg, Macao, Malaysia, Malta, Martinique, Mexico, Namibia, Netherlands, New Caledonia, New Zealand, North Macedonia, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Republic Of Tanzania, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, Viet Nam.

Official Source

View FDA Recall →