Normed Charcot Osteotome Chisel (Various sizes)
FDA Recall Z-1360-2018 · Posted None
Advertisement
Recalling FirmZimmer GmbH
Recall StatusTerminated
Product CodeKDG
Date Initiated2018-03-15
Date Terminated2019-04-25
Reason for Recall
Product may contain a gap between the handle and metal chisel which can lead to failure to properly disinfect or sterilize between use.
Firm Action
On March 15, 2018 an URGENT MEDICAL DEVICE RECALL LETTER was issued to affected hospitals with the following instructions:
1. Review this notification for awareness of the contents.
2. Assist your Zimmer Biomet sales representative to quarantine immediately all affected instruments.
3. Your Zimmer Biomet sales representative will remove the affected instruments from your facility
Questions or concerns can be directed to a zimmer biomet representative.
Root Cause
Under Investigation by firm
Distribution Pattern
CO, IN, TX, IL
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement