ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003
FDA Recall Z-1363-2018 · Posted None
Advertisement
Recalling FirmOrthofix, Inc
Recall StatusTerminated
Product CodeNKG
Date Initiated2018-02-23
Date Terminated2019-06-18
Product Quantity397 units
Reason for Recall
Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or
(2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.
Firm Action
The firm disseminated the recall notices on 02/23/2018 by mail. The firm requested the return of the screws.
Root Cause
Nonconforming Material/Component
Distribution Pattern
US
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.