Terminated

ADVIA Chemistry Lipase, Ref. No. 01984894 B01-4840-01, Siemens Material No. 10311896, UDI 00630414482446 - Product Usage: Lipase measurements are used in the diagnosis and treatment of diseases of the pancreas, such as acute pancreatitis and obstruction of the pancreatic duct.

FDA Recall Z-1363-2020 · Posted 2020-01-30

Recall StatusTerminated
Product CodeCHI
Date Initiated2019-12-20
Date Terminated2020-12-21
Product Quantity3224

Reason for Recall

Firm has received customer complaints regarding failed calibrations and increased imprecision of quality control and patient samples when using this product.

Firm Action

On December 20, 2019, the recalling firm distributed Urgent Medical Device Recall letters to affected customers. Customers were provided with instructions for determining whether the Lipase reagent cartons in their inventory were impacted. The next reagent lot is expected to be available by February 2020. Customers were asked to review the letter with their Medical Director, review their inventory for affected products, and assess their laboratory's replacement needs as applicable. A customer response form was included with the recall communication so that customers may confirm that they have received the notification as well as report the quantity of replacement product required.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CT, FL, GA, IA, KS, MD, MI, NJ, NM, NY, OH, OK, PA, TN, TX, WA.